Product Description

Peristaltic Pump Footswitch

Peristaltic Pump Footswitch

Suitable Drive
Type Work Mode Operational Steps
JK-1 a) Gated: Pump stops when it is powered on. Pump runs when press down the footswitch; and pump stops when release the footswitch.
b) Trigger: Pump stops when it is powered on. Pump runs when footswitch is pressed once, and pump will stop when footswitch is pressed again.
a) Insert DB9 plug into external port on the rear of peristaltic pump.
b) Pump doesn’t run when it is powered on, then using footswitch to control the pump.
JK-2 only one work mode a) Insert DB9 plug into external port on the rear of FK-1C.
b) Power on FK-1C Dispensing Controller, turn internal/external control switch to ‘external control’.
c) Press footswitch once (Note: press one time is enough), FK-1C will run until the setting time run out.
JK-3 only one work mode a) Insert DB15 plug into external port in the rear of the peristaltic pumps.
b) The initial state of the peristaltic pump stopped when power on it.
c) Pump runs when press down the footswitch; and pump stops when release the footswitch.
JK-4  a) Gated: Pump stops when it is powered on. Pump runs when press down the footswitch; and pump stops when release the footswitch.
b) Trigger: Pump stops when it is powered on. Pump runs when footswitch is pressed once, and pump will stop when footswitch is pressed again.
a) Insert DB9 plug into the external port in the rear of syringe pumps.
b) Power on syringe pump, syringe pump should restore the previous parameters before power-down then footswitch can be used to control the pump.
JK-5 a) Trigger mode: Trigger: Pump stops when it is powered on. Pump runs when footswitch is pressed once (press down and release), and pump will stop when footswitch is pressed again.  a) Insert DB15 plug into external port in the rear of the pump.
b) Power on syringe pump, syringe pump should restore the previous parameters before power-down, then footswitch can be used to control the pump.
JK-6 a) Trigger mode: Trigger: Pump stops when it is powered on. Pump runs when footswitch is pressed once (press down and release), and pump will stop when footswitch is pressed again.  a) Connect the wire of footswitch to the corresponding ends in the external control port of pump.
b) Power on syringe pump, syringe pump should restore the previous parameters before power-down then footswitch can be used to control the pump.